16
Input Configuration Menu
60
16.2 Ad.RAtE Enter ANALOG TO DIGITAL RATE SUBMENU:
Press ‘RESET/ENTER 7) Display flashes "0" or previous selection.
Press ‘/MAX 8) Scroll through the 6 available selections as 0, 1, 2, 3, 4, and 5 for
your choice of selection. Refer to below Table for your specific application
conversion speed. Only Rate 0 and 1 can reject 60 Hz / 50 Hz noise.
Refer to Table 33-1 in Specification Section for different input and outputs types.
Rate 0 1 2 3 4 5
Time for 60hz(ms) 134.9 68.2 34.9 18.2 9.9 5.7
Time for 50Hz(ms) 161.9 81.9 41.9 21.9 11.9 6.9
Table 16-1. A/D (Analog to Digital) Rate Option Table
Press ‘RESET/ENTER 9) Display advances to "MOdE" Input Mode Configuration Submenu. If a
change was made, the controller will also show "StOREd" stored
message momentarily.
16.3 MOdE Enter INPUT MODE CONFIGURATION SUBMENU:
This submenu is only accessible if "tC" Thermocouple or "bRIdGE"
Bridge is your selection of input type.